User's Manual
SPECIFICATIONS
Ultraview
Digital
Telemetry
90343,
90347,
90478-Q,
90478-T,
90478-V,
90479-A
SpO
2
SpO
2
Sensor Interface —
Red LED drive (max): 175 mA peak at 10%
duty cycle
IR LED drive (max): 105 mA peak at 10% duty
cycle
SpO
2
Measurement Method — Functional
saturation (oxygen saturation of functional
hemoglobin)
SpO
2
Measurement Mode —
Continuous, episodic (2 minutes, 5 minutes,
and 30 minutes) sampling intervals; factory
default setting is continuous
ST Segment Analysis
Resolution — 0.08 mm
Range — ±9 mm (1 mV = 10 mm)
Leads — ST Segment Analysis continously
performed on up to seven leads
Alarms — Single lead or multiple leads; individual
leads can be deselected
Trends — Up to 24 hours of trend data can be
displayed in 1.5-, 3-, 6-, 12-, or 24-hour time
tracks
DISPLAY CHARACTERISTICS
ECG Display
Heart Rate Range — 30 to 300 bpm; heart rates
> 300 bpm are displayed as “+++”
Heart Rate Alarm Limits — High: 5 to 300 bpm,
Low: 0 to 200 bpm; alarms automatically
enabled over a range of 40 (adult) or 100
(neonatal) to 300 bpm
Accuracy — ±1% or ±2 beats per minute
(whichever is greater)
Numeric Update Rate — Every 3 seconds or
immediately at the onset of an alarm
Trace Sweep Speeds — 50, 25, 12.5 mm/sec
SpO
2
Display
Measurement Range — 30 to 100% O
2
Saturation
Saturation Accuracy: — Sensor Dependent
Saturation Resolution — ±1%
Pulse Rate Range — 30 to 250 bpm
Pulse Rate Resolution — 1 bpm
Alarms — High and low saturation values; factory
default limits are: high 100%; low 85%
High range: 31% – 100%
Low range: 30% – 99%
Numeric Update Rate — Every 2 seconds for
continuous SpO
2
readings
NIBP Display
(Refer to specifications for the 90217 ABP
Monitor)
Measurement Range (adult only) —
Systolic: 8.0 – 35.0 kPa (60 – 260 mmHg)
Diastolic: 9.0 – 27.0 kPa (30 – 200 mmHg)
Mean: 5.3 – 31.0 kPa (40 – 230 mmHg)
Pressure Accuracy — ±2% or ±3 mmHg
(whichever is greater)
Resolution — 1 mmHg
Time Between Readings — selectable, from 6 to
120 minutes
Alarms — High and low alarms for all measured
parameters
High range: 8.0 – 35.0 kPa
(60 – 260 mmHg)
Low range: 4.0 – 27.0 kPa
(30 – 200 mmHg)
TRANSMITTERS (90343, 90347)
ECG Transmission — View 2 of 7 available
leads from four vectors (90343, 90347)
synchronized RF digital signal
Electrode Configuration — Individually
replaceable DIN standard safety lead wires
Multi-parameter Transmission (90343) — SpO
2
(saturation, SpO
2
sensor status, pulse rate)
and optional NIBP (systolic, diastolic, mean
pressure, measurement time, alarm
conditions) with the model 90217 ABP monitor
Additional Data Transmitted — Patient record,
low battery indicator, pacer flag, patient ID
code, and electrode connection status
Output Power — 4 mW ERP, typical
Spectral Efficiency — 0.11 bps/Hz
External Indicator — Yellow LED flashes when
battery level is low
Battery — 9 V battery; refer to Table 1 for battery
life expectancy
Frequency Band Option —
-04: 602-608 MHz
-05: 608-614 MHz
-06: 614-620 MHz
TRANSMITTER PHYSICAL DIMENSIONS
90343 (Multi-Parameter)
Height: 5.25 in (13.3 cm)
Width: 2.85 in (7.2 cm)
Depth: 1.18 in (2.9 cm)
Weight (w/out battery): 8.5 oz (241.0 gm)
90347 (ECG-only)
Height: 5.25 in (13.3 cm)
Width: 2.85 in (7.2 cm)
Depth: 0.98 in (2.5 cm)
Weight (w/out battery): 6.78 oz (192.7 gm)