User's Manual
SPECIFICATIONS
PARAMETER CHARACTERISTICS
ECG
Maximum Input — ±4 mV ( ±10%)
DC Offset — Up to ±300 mV, with no more than
2% signal amplitude degradation
Overdrive Recovery Time — < 1 second circuit
settling time with offset voltage < 500 mV
Noise — < 30 µV p-p, refered to input (rti), at 30
Hz bandwidth
CMRR — > 85 dB (monitor mode)
QRS Detection — Detects QRS complexes with
amplitudes of 0.2 to 4.0 mV (adult) or 0.15 to
4.0 mV (neonatal)
Defibrillator Protection — Meets IEC
60601-2-27, AAMI EC-13
Resolution — 2.5 µV per LSB, rti
Input Impedance — > 10 MΩ minimum
differential at 10 Hz
Gain Accuracy — ±5 %
Pacer Rejection — Baseline shift < 0.2 mV
(measured at ECG x 1,000 output)
Pacer Detection — Detects pacer pulses of ±2
mV to ±700 mV with pulse widths of 0.2 to 2
msec and rise times 10% of width not to
exceed 100 µsec
Signal Bandwidth — 0.05 to 30 Hz ±10% (-3
dB)
Sample Rate — 120 samples per second
Ultraview
™
Digital
Telemetry
90343, 90347,
90478-Q,-T,-V,
90479-A
■ 602-620 MHz frequency operation,
including 608-614 MHz Wireless Medical
Telemetry Service (WMTS) band
■ Touchscreen control of all module
functions and compatible with all
Ultraview Care Network
™
monitors
■ Multi-parameter telemetry monitoring
ECG, SpO
2
, and NIBP (optional)
■ Lightweight, water resistant transmitters
■ Diversity antenna system
■ Tunable modular receiver
■ Modular receiver converts bedside
monitors to telemetry operation
■ Multi-lead ECG with ST segment analysis
option; comprehensive arrhythmia and
ST trending
■ Module Configuration Manager enables
the hospital to customize the receiver’s
ECG patient monitoring functions to
specific patient populations, clinical
protocols, or operating preferences
■ Graded alarm functions enables the
hospital to define high, medium, or low
alarm tones according to critical,
warning, or advisory event severity
Diversity Antenna
System
90217 ABP Monitor
90478 Modular Receiver
90479-A Modular
Receiver Housing
90343
Digital
Telemetry
Multi-parameter
Transmitter
90347
Digital
Telemetry
ECG
Transmitter
(options Q, T, and V)
mmHg
mmHg
model 90217
• Install three, fresh “AA” cells
( 1.5 Alkaline or NiCad).
Observe polarities.