Figure 22-10: Split screen, central with numeric key display
Agent Parameter Label View
When in AUTO ID mode, the inspired and expired agent labels are displayed in
normal video, and in the MANUAL mode, the labels are displayed in reverse
video.
In MANUAL mode, full-screen large text view, the mixed condition is reported by
displaying MIX under the parameter label area next to the units of measure. In the
full-screen small text view, the parameter labels alternate between the mix label
and agent type label. In AUTO ID mode, the agent labels are replaced by “I MIX”
and “E MIX”.
No Breath Mode View
Normal mode of operation is in effect as long as breathing is detected. Twenty
seconds after the last breath is detected, the analyzer switches to “no breath
mode” automatically and the EtCO
2
, Resp rate, expired agent, FeO
2
, and expired
N
2
O numeric values are shown as ???. The momentary data is displayed in place
of the inspired numeric values.
Inspired nitrous oxide data Fractional expired O
2
Inspired agent Expired agent
Fractional inspired O
2
쐅 Expired nitrous oxide data
Respiration rate in breaths/minute 쐈 Inspired carbon dioxide data
Alarms ON bell (refer to Table 1 on page 22-15) 쐉 Apnea alarm setting (the delay value)
End tidal carbon dioxide data 씈 Bed ID and patient name (remote displays only)