5. Place the optical connector into the optical module with the word TOP facing
up. Close the module (refer to Figure 20-3).
Figure 20-3: Placing the optical connector into the optical module
6. With the inner wrap still in place, press down on the black optical reference at
the point labeled PUSH. Listen for a click.
The catheter is now ready for pre-insertion calibration. After you complete the
calibration, prepare the patient for catheter insertion.
Performing Pre-insertion (in vitro)
Calibration
This calibration procedure tales up to 72-seconds. When calibration is successful,
the message PRE-INS CAL COMPLETE displays and the event mark P displays
on the oximetry trend.
If the catheter fails calibration, verify the following:
A secure cable connection exists between the optical and Sv
O
2
modules.
The catheter tip is inserted properly within the optical connector.
You have pressed the optical reference with enough force to hear the click,
indicating a secure connection.
You tested the catheter by replacing it with a known good catheter to ensure
proper catheter function.
Repeated calibration failure with a known good catheter may indicate a defective
optical module. Replace the optical module if necessary.
To initiate pre-insertion
calibration:
1 Touch SVO2.
2 Touch CALIBRATE.
3 Touch PRE-INSERTION.
4 Touch YES.