Diagnostic Messages
Message Cause/Action
MONITOR SUPPORTS xx PARAMETERS.
ADDITIONAL PARAMETERS MAY OVERLOAD IT
Data for more than xx parameters is being input to the monitor.
System performance may decline. Unplug modules to improve
performance.
UNABLE TO SUPPORT A NEW MODULE AT THIS
TIME
There is not enough system memory to support a new module.
Unplug modules to free memory.
UNABLE TO SUPPORT A NEW CHANNEL AT
THIS TIME
There is not enough system memory available to support a new
channel. Detach channels to free memory.
DIAGNOSTIC ERROR ENCOUNTERED LOADING
MODULE
Checksum error detected loading module. Module is unsupported.
DIAGNOSTIC ERROR ENCOUNTERED LOADING
CHANNEL
Checksum error detected loading channel. Channel is unsupported.
OUT OF MEMORY - DELETING HEMO ENTRY
There is not enough system memory available to the application to
complete the requested operation. Try again later.
OUT OF MEMORY - DELETING RESP ENTRY
There is not enough system memory available to the application to
complete the requested operation. Try again later.
OUT OF MEMORY - DELETING OXY ENTRY
There is not enough system memory available to the application to
complete the requested operation. Try again later.
OUT OF MEMORY - DELETING RENAL ENTRY
There is not enough system memory available to the application to
complete the requested operation. Try again later.
CREATE NEW ENTRY FAILURE - MAXIMUM NUM
ENTRIES EXCEEDED
The maximum number of Calcs entries are in use. Delete existing
entries to create new ones.
REMOTE MONITOR NOT RESPONDING
Calcs timeout expired waiting for remote GDS response. Verify
remote monitor is on network, and try again.
ERROR READING STORED CALCS DATA
Bad return code from GDS on data read. GDS is possibly corrupt.
Reboot monitor.
ERROR STORING CALCS DATA
Bad return code from GDS on data store. GDS is possibly corrupt.
Reboot monitor.
ERROR DELETING STORED CALCS DATA
Bad return code from GDS on data delete. GDS is possibly corrupt.
Reboot monitor.
NO TRENDS AVAILABLE FOR THIS BED There is no trend data in GDS. Wait one minute and try again.
NO OTHER BED ON THE NETWORK HAS ACTIVE
PARAMETERS
This message is in response to a remote request. The monitor has
determined that there are no beds on the network with active
parameters so remote operations are not possible at this time.
THERE ARE NO PARAMETERS ACTIVE ON THIS
BED
The selected bed has no active parameters. In order to remotely
work with the monitor at the selected bed, it must have active
parameters.
PARAMETER IS NOT AVAILABLE TO MONITOR Failed attempt to attach to channel. Try again.
COMMUNICATIONS WITH REMOTE MONITOR
INTERRUPTED
Lost connection to remote monitor. Verify that the remote monitor is
on the network and retry remote operation.
THIS OPERATION WILL CAUSE THE MONITOR
TO RESET
The requested operation will reboot the monitor. If you do not want
this to occur, touch PREVIOUS MENU or NORMAL SCREEN.
THIS INSERTION WILL CAUSE PRIOR
ASSIGNMENT TO BE LOST
The requested zone assignment will cause a previously assigned
zone to be lost. Touch PREVIOUS MENU or NORMAL SCREEN to
abort the operation.