User's Manual
Ultraview Care Network
20-4
Setting Up SvO
2
Monitoring
In addition to the Spacelabs Medical module, venous oximetry monitoring requires
an Abbott Oximetrix 3 optical module with connecting cable and an Abbott
Opticath catheter. You must correctly connect all necessary cables, prepare the
catheter and patient for this invasive procedure, and calibrate the catheter to
begin monitoring.
The system provides two calibration methods. The patient’s current status
determines the proper calibration method. You can calibrate the catheter:
• while in its sterile package, or
• after it has been placed within the pulmonary artery, and you can verify
adequate light intensity during monitoring.
Refer to Preparing the Catheter on page 20-5 and Performing Pre-insertion (in
vitro) Calibration on page 20-6 for more details.
Display Detail
Figure 20-1 provides a trend of oximetry values and the current light intensity
value.
Figure 20-1: Typical venous oximetry display
Current time base — 1HR
Venous oximetry trend graph
Event marks — I (in vivo calibration initiated), C (in vivo calibration complete)
Light intensity display marks (vertical lines) superimposed over trend graph
Light intensity bar graph (light meter)
Real time intensity signal level (two vertical lines)
CAUTION:
• The Spacelabs Medical Sv
O
2
module will not correctly
operate with any other catheter for Sv
O
2
monitoring.
To set up SvO
2
monitoring:
1 Place the catheter’s optical
connector into the optical
module on the connecting
cable.
2 Insert the connecting cable into
the SvO
2
module.