User's Manual

29-7
Patient Data Logger Troubleshooting Guide
Clinical Situation Possible Causes Solution
Data is displayed with
improper spacing or double
spacing
The PDL interface sends a
carriage return/line feed
sequence at each end-of-line.
The external device may not be
set up properly.
Set the external device for “0” line
feed.
No data is displayed or
printed on the external
device
There is a power problem or the
cables are faulty. The device may
not be set up properly.
Check the power and cables. Ensure
that the device is in the online mode
and that RS-232 port requirements
are satisfied.
Check for RS-232 compatibility at the
monitor and at the external device.
Check the monitor port assignments
and port connections.
Data is lost or garbled The cable is faulty. Check the cables.
The parity is set incorrectly. Verify the baud rates and parity
settings.
Baud rate settings may be
inappropriate.