Trends Troubleshooting Guide
Clinical Situation Possible Cause Solution
Current patient data not
being added to trends
If alarms are suspended data may not
be trended at the bedside monitor.
Turn ON alarms or have your system
administrator enable your system to
trend while alarms are suspended.
EXPAND key ON
(trend data will not be lost).
Change EXPAND to OFF.
Resume ECG processing.
Cursor not in home position
(trend data will not be lost).
Move the cursor to the extreme right
position.
PCWP trend not
available
PCWP values not saved. Save the PCWP values.
Incorrect unit of
measure displayed
Incorrect unit of measure configured
for system.
Contact your system administrator.
CALCS data not
trending
Trends cannot trend CALCS. Use the CALCS menu to display this
data.
NO TRENDS
AVAILABLE FOR THIS
BED message is
displayed
No trend data has yet accumulated for
the selected parameter.
Allow sufficient time for data to
accumulate.
Numerous entries with
??? instead of vital
signs
Alarm Suspend was selected
frequently.
Contact your system administrator to
collect data during Alarm Suspend
periods.
ECG/Resp processing suspended. Do not suspend ECG processing unless
the patient is off the system.
Ensure that ECG/Resp amplitude and
signal quality are sufficient.