User's Manual

Ultraview Care Network
24-12
Storing a Setup
Once you have set the EEG module’s parameters the way you want them, you
may store the current settings to one of six setup keys. Access the Initial Setup
menu by first touching the SETUP key from the Main menu, then touching INITIAL
SETUP. From the Initial Setup menu, touch STORE SETUP. You will then see six
setup keys labeled SETUP #1, SETUP #2, etc. Store the current settings by
touching one of the setup keys.
Recalling a Setup
To recall a setup, make sure processing is stopped, then touch the SETUP key.
Touch RECALL SETUP. You will then see the retain or erase archived EEG data
keys if data is stored in the module. If you want to change your setup, you will
have to erase any archived EEG data. If you choose to retain archived EEG data,
you cannot select another setup.
EEG Electrode Lead Selection (Montage)
To select which leads you have placed and their placement sites, touch the
MONTAGE key (under SYSTEM PARAMS in the Initial Setup) to enter a sub-
menu in which the electrodes can be selected and the impedance of each
electrode can be measured. First, select either two or four channels for EEG
recording. If two channel recording is selected, the CH3 and CH4 keys will be
inactive, and the electrode names are removed.
The following procedure is used to select electrode names for each channel:
1. Touch the desired channel key (CH1, CH2, CH3, CH4).
2. Touch the CLEAR CHANNEL key to delete previously defined electrodes for
the channel.
3. Touch one of the 25 keys in the montage to select the first electrode. The key
will flash on and off, and appear in the channel key as the positive input. Select
a second electrode. The key will again flash on and off, and your selection will
appear in the channel key as the negative or reference input.
!
Processing must be stopped before a setup can be stored.
The message EEG Setup Successfully Stored is displayed on
the message line when the setup is stored.
!
The first electrode name displayed in the channel key is the
positive input, while the second electrode name is the negative
or reference input.
When selecting the next channel, the electrodes that were
selected for the first channel will stop flashing, but will remain
highlighted.
To store a setup:
1 Touch EEG.
2 Touch SETUP.
3 Touch INITIAL SETUP.
4 Change desired parameters.
5 Touch STORE SETUP
6 Touch the setup key to which
you want to store the current
setup.
To recall a setup: (processing
must be off)
1 Touch EEG.
2 Touch SETUP.
3 Touch RECALL SETUP.
4 Touch ERASE ARCHIVED
EEG DATA (if data is stored in
module).
5 Touch RECALL SETUP
6 Select desired setup
To select EEG electrode leads:
1 Touch EEG.
2 Touch SETUP.
3 Touch INITIAL SETUP
4 Touch SYSTEM PARAMS.
5 Touch MONTAGE.
6 Touch ERASE ARCHIVED
EEG DATA (if data is stored in
the module).
7 Select 2 or 4 channel.
8 Select desired channel.
9 Touch CLEAR CHANNEL.
10 Select desired electrode
placement sites.
11 Repeat steps 8 through 10 until
all channels are defined.