User's Manual

Ultraview Care Network
15-14
Bradycardia Alert
Type 1
The heart rate is divided into eight groups of 20-seconds each. Group H
represents the current 20-seconds.
An alert will activate if Groups G and H are below 90 bpm.
Figure 15-3: Type 1 bradycardia alert
CAUTION:
Any time the alert activates, the heart rate must return to
normal before the Alert can be reactivated. Reset the alert
by pressing the F-ALERT RESET key on the screen.
If a group of heart rates is less than 30 bpm, the algorithm
considers this condition a loss of signal and a NO ALERT
will occur.
!
All representations of fetal heart rates in this chapter are
computer generated.