User's Manual
Arrhythmia
9-9
Display Detail
The system stores and displays arrhythmia review data in two categories:
non-template forming classes and template-forming classes.
The non-template forming class displays a waveform that is 6.25-seconds long
and is representative of the most recent occurrence.
The template-forming class displays two waveforms. The waveform on the left is
1.25-seconds long; the arrhythmia that originated the class is centered. The
waveform on the right is 4.5-seconds in length and the most recent occurrence of
the class is centered.
The lead, frequency, and time and date of last occurrence appear to the right of
each presentation.
Figure 9-1: Arrhythmia waveforms
Non-template Forming Class Example
Template Forming Class Example
Lead II
Last MINUTE: 1
Last HOUR: 1
Last
OCCURRENCE:
04:44AM
5 Jun 2003
Lead VI
Last MINUTE: 2
Last HOUR: 6
Last
OCCURRENCE:
04:40PM
5 Jun 2003
(Tachycardia)
(Couplet)