User's Manual

SpO
2
14-11
ADAPTER DISCONNECTED — CHECK CONNECTION AT MODULE
The module does not detect an adapter cable connected to the front panel.
Check proper adapter cable connection.
If the message persists and the adapter cable is secure, replace the adapter
cable.
The channel will drop and the alarm will stop after approximately 10-seconds.
On remote view, there may be no audible alarm on the remote monitor before
the channel drops.
SENSOR DISCONNECTED — CHECK CONNECTION AT ADAPTER CABLE
The module does not detect a sensor. Check the proper connection of the
sensor with adapter cable.
If the message persists, replace the sensor and/or the adapter cable.
SENSOR OFF PATIENT — CHECK CONNECTION AT PATIENT
The module does not detect a valid sensor input signal. Check the patient for
proper sensor placement.
The tissue between the LED and photodiode is too transmissive. If the sensor
placement seems correct and the message persists, try a sensor site with a
thicker tissue bed.
FAULTY SENSOR — REPLACE SENSOR
The LED and/or photodiode have failed. Replace the sensor.
AMBIENT LIGHT INTERFERENCE — COVER SENSOR AREA
The sensor is receiving external light interference from a bright light source
near the sensor. Shield the sensor from the external light source.
The sensor photodiode and LEDs are misaligned on flexible sensors thereby
allowing light to enter. Realign the sensor photodiode with LEDs.
If the message persists, replace the sensor.
INSUFFICIENT SIGNAL — REPOSITION OR REPLACE SENSOR
Insufficient signal for proper operation as evidenced by a low deflection on
Sensorwatch signal strength bar.
Poor sensor application or site. Correctly re-apply or reposition to a better
perfused site or massage site.
If the message persists, replace the sensor.
!
This message is not available with disposable SpO
2
sensors or
non-clip type sensors.