User's Manual
SpO
2
14-9
Restoring User-Defined Settings
With the Restore Settings feature, you can restore the user-configurable settings
to the defaults previously stored as user settings. User-configurable settings are
listed in Setting User-Defined Default Values on page 7-7.
Suspending/Resuming SpO
2
Processing
When you touch YES in the Suspend Processing menu, analysis and display of
the SpO
2
data is suspended. If you touch NO, the display returns to the SpO
2
menu without affecting processing.
When you suspend SpO
2
processing:
• The message SPO2 PROCESSING SUSPENDED appears in the SpO
2
waveform zone.
• Question marks replace the SpO
2
and pulse rate values.
• The message SPO2 ALM OFF replaces the alarm limits.
• The keys in the SpO
2
Alarm Limits menu display outline to indicate they are
disabled.
• The SUSPEND PROCESSING key changes to RESUME PROCESSING.
Recording SpO
2
Waveforms
You can print SpO
2
waveforms and values. Refer to Printing on page 4-7 for
additional information.
!
• RESTORE SETTINGS changes the user-configurable settings
for all parameters in the module.
• Once restored, alarm ON/OFF status and alarm limits are reset
to the previously stored user settings.
!
• This feature is not supported in telemetry products.
To restore user-defined settings:
Configurations with ECG
1 Touch ECG.
2 Touch SETUP.
3 Touch RESTORE SETTINGS.
4 Touch YES.
To restore user-defined settings:
Configurations without ECG
1 Touch TEMP.
2 Touch RESTORE SETTINGS.
3 Touch YES.
To suspend SpO
2
processing:
1 Touch SPO2.
2 Touch SUSPEND
PROCESSING.
3 Touch YES.
To resume SpO
2
processing:
1 Touch SPO2.
2 Touch RESUME
PROCESSING.
3 Touch YES.
To print SpO
2
waveforms:
1 Touch RECORD.
2 Touch flashing SPO2
parameter key.