User's Manual

13-12
NIBP Troubleshooting Guide
Clinical Situation Possible Cause Solution
No NIBP screen key is
displayed
Module not inserted correctly. Remove and re-insert the module.
No NIBP readings can be
obtained
Incorrect or inoperative cuff in use. Replace with the cuff known to be
operative.
Cuff tubing is attached to an adult
outlet, but monitor is configured in
the neonatal mode (or vice versa).
Connect tubing to the correct outlet.
Correlate monitor mode, cuff and
patient type.
Tubing is kinked. Locate kink and straighten tubing.
Some arrhythmias (for example,
atrial fibrillation and frequent
ventricular ectopy) may cause a
single or repeated failure to obtain
a reading (may be due to true
beat-to-beat variations in
pressure).
Document arrhythmia if present, verify
pressure with another method, then
follow hospital procedure for care of
this type of patient.
Excessive patient motion or
muscle contractions associated
with shivering or severe pain.
Ensure that patient is quiet with
minimal movement during NIBP
readings. Minimize the patient’s
shivering.
Blood pressure outside of
measurement range.
Verify extremely high or low pressures
with another method.
Intermittent or complete
failure to operate
Hardware error (codes 10, 20, and
30) detected during previous
measurement.
Check the presence of the RESET
NIBP key in the CHANGE CONFIG
menu. Touch RESET NIBP to
re-enable monitoring.
Remove the module from service and
call a qualified service person if this
condition occurs repeatedly.
Apparent incorrect value Wrong size cuff for patient. Measure patient’s limbs at the
midpoint. Match limb measurement to
the range specified on the cuff
(undersizing the cuff results in the
greatest degree of error).
Cuff is damaged. Replace with a good cuff.
Excessive patient motion,
shivering or severe pain.
Ensure the patient is quiet with minimal
movement during NIBP readings.
Minimize the patient’s shivering.
False high readings may be the
result of venous congestion
caused by frequent readings.
Reduce the frequency of the readings.
Cuff too loose or positioned
incorrectly.
Tighten the cuff or reposition it
appropriately.