User's Manual

12-14
Respiration Troubleshooting Guide
Clinical Situation Possible Cause Solution
Inaccurate respiratory rate
or 0 displayed. Question
marks displayed instead
of rate
Respiration too shallow for normal
detection.
Touch the SHALLOW/NORMAL key to
highlight SHALLOW.
ECG electrode contact or
placement poor.
Apply new electrodes. Make sure to
properly prepare the skin; position
electrodes on the chest where the
chest expansion is the greatest.
Incorrect lead selection for
respiration.
Select the appropriate lead. Best lead
selection is typically RA-LA for adults
and RA-LL for infants.
CVA artifact. Assess the patient for apnea. Reselect
lead for better signal quality.
No respiration waveform.
LOSS OF SIGNAL
message is displayed
ECG electrodes or patient cable not
attached.
Select another lead.
Reconnect the leads or the patient
cable.
No respiration waveform
is displayed
Channels have not been configured
to display respiration.
Select RESP ON in the ECG Display
Format menu.