User's Manual

Ultraview Care Network
10-8
Reviewing ST Data
Use the ST Review menu to display and review ST segment data. A snapshot of
ST segments for all available leads is displayed. The time and date of the ST
snapshot is displayed on the menu prompt line. Touch the PRIOR SET or NEXT
SET key to display ST data at other time points. Figure 10-5 is an example of a ST
segment snapshot.
Nine ST snapshots can be stored for display. The oldest ST snapshot, not marked
as SAVED, will be deleted to make room to store a new snapshot. To save an ST
snapshot, touch the SAVE SET key.
ST snapshots are automatically stored whenever an ST alarm occurs or at pre-
selected time intervals. To store the ST snapshots at periodic intervals, select
TIME SAVE YES. Available intervals are 5-, 10-, 15-, 30- or 60-minutes (only
available in the Module Configuration Manager).
The factory default setting for the time save interval is 15-minutes.
Figure 10-5: ST segment display
Clearing ST Data
Touching the CLEAR key and then confirming your choice by selecting YES
clears the currently displayed data for all leads from memory. Data is also cleared
from the trends.
To review ST data:
1 Touch ECG.
2 Touch REVIEW.
3 Touch ST REVIEW.
AVR
AVF
V4
V6
V1
V2
V3
III
II
I
-0.16 mm
-0.16 mm
0.16 mm
2.88 mm
4.32 mm
0.16 mm
0.16 mm
0.16 mm
0.00 mm
1.44 mm
AVL
0.24 mm
To clear the displayed ST data:
1 Touch ECG.
2 Touch REVIEW.
3 Touch ST REVIEW.
4 Touch CLEAR.
5 Select YES.