User's Manual

Pressure
18-9
Selecting a Filter Frequency
When noise and other interference appears on the pressure waveform, you can
adjust the filter frequency to minimize its effect.
A higher filter frequency shows greater detail, but can also produce more artifact.
A lower filter frequency smooths the waveform and may help diagnose transducer
or catheter problems such as under damping or ringing.
The filter frequency may be set within the range of 3 to 40 Hz.
Restoring User-Defined Settings
With the Restore Settings feature, you can restore the user-configurable settings
to the defaults previously stored as user settings. User-configurable settings are
listed in Setting User-Defined Default Values on page 7-7.
!
RESTORE SETTINGS changes the user-configurable settings
for all parameters.
Once restored, alarm ON/OFF status and alarm limits are reset
to the previously stored user settings.
To adjust the filter frequency:
1 Touch the desired pressure
parameter key.
2 Touch SETUP.
3 Touch FILTER.
4 Touch the arrow keys to choose
the desired setting.
To restore user-defined settings:
Configurations with ECG
1 Touch ECG.
2 Touch SETUP.
3 Touch RESTORE SETTINGS.
4 Select YES.
To restore user-defined settings:
Configurations without ECG
1 Touch TEMP.
2 Touch RESTORE SETTINGS.
3 Select YES.