User's Manual

Arrhythmia
9-13
Excluding Classes from Trends
Use this function to exclude specific classes from the group trend for that class as
well as from the dominant trend graph (only valid with CPL and ABN classes).
This function defaults to YES indicating that all classes will be included. When you
select a class, the template for that class displays on the screen. It is followed by a
trend graph that plots all of the occurrences of that class over the selected trend
graph timebase.
Positioning the Cursors
The cursors are small, bright lines that move along the bottom of the trend display
and allow you to view the number of trended events that occurred between any
two points in time displayed on the trend graph.
Initially, the left (L) cursor is located flush with the left edge of the trend graph, and
the right (R) cursor is located flush with the right edge of the trend graph. The
number of trended events between, and inclusive of, the cursor points appear
below the trend graph. If the cursors move past each other, the L cursor becomes
the R cursor and vice versa. The highlighted L or R portion of the CURSOR key
changes accordingly.
Printing Arrhythmia Data
All printouts of ECG or arrhythmia data are annotated with the following data:
Bed identification
Time and date of the printout
Lead designator
You can print all arrhythmia data for all classes or individually selected classes.
You can also print individually selected arrhythmia trend plots.
Selecting PRINT ALL prints all ST events and all arrhythmia classes. PRINT ALL
also prints all leads.
To exclude a class in a group
trend:
1 Touch ECG.
2 Touch REVIEW.
3 Touch ARRHYTHMIA
REVIEW.
4 Select an arrhythmia class.
5 Touch CLASS TREND.
6 Select INCLUDE NO.
To position the cursors on the
trend graph:
1 Touch the CURSOR to highlight
either the L or R portion.
2 Touch the trend plot to position
the cursor near the desired
point.
3 Use the arrow keys for
adjustment of the cursor.
To print recordings of ALL
arrhythmia events:
1 Touch ECG.
2 Touch PRINT.
-OR-
3 Touch PRINT ALL.
4 Touch ARR CLASSES.
To print selected arrhythmia
classes:
1 Touch ECG.
2 Touch REVIEW.
3 Touch ARRHYTHMIA
REVIEW.
4 Select an arrhythmia class.
5 Touch PRINT.
To print selected arrhythmia
trends:
1 Touch ECG.
2 Touch REVIEW.
3 Touch ARRHYTHMIA
REVIEW.
4 Select an arrhythmia class.
5 Select CLASS TREND or
GROUP TREND.
6 Touch PRINT.