User's Manual
Fetal Monitoring
15-29
Recharging Batteries
1. The battery packs are enclosed in the top half of the Mermaid ultrasound and
Toco transducers (Figure 15-17). To remove the packs (a) from the transducer
elements, twist the pack counterclockwise. There will be a noticeable
movement. Simply pull the battery pack to disengage.
2. Insert a battery pack (a) in the docking station (e) on the recharger by aligning
the ridge with the red dot. The battery pack should fall into the docking station
with a noticeable movement.
Figure 15-17: Wireless transducer Mermaid system
3. Twist the battery pack clockwise slightly to lock it into the docking station. The
bottom of the transducers do not need to be charged and should be stored
along with the transmit badge in the compartment located in the rear of the
recharger unit (f).
!
• Thoroughly dry the battery pack before inserting it into the
charger.
• The Mermaid battery charger may be used in the patient vicinity
if it is powered by one of the following certified, medical-grade,
power supplies:
- Ault, Model SW172; Rated Input: 100-240VAC, 0.75-0.35A,
50/60Hz; Rated Output: 12VDC, 2.75A (P/N 010-1554-00)
- Ault, Model VM100AMB1203B01; Rated Input: 100-250VAC,
0.3A, 50/60Hz; Rated Output: 12VDC, 0.85A (P/N 1043900-
108)
• These power supplies are designated Class I by the applicable
safety standards. Medical-grade, grounded outlets and power
cords are required to ensure patient safety margins.
• Three lights indicate a full charge and 8-hours (typical) of
operation. Two lights indicate that 2-hours (minimum) of
monitoring time is left. One light indicates that less than 2-hours
of monitoring time is left. When the one light starts to flash, less
than 15-minutes of battery capacity is left.