User's Manual

Fetal Monitoring
15-25
Parameter Configurations
Table 2: Fetal Parameter Configuration
Parameter
Available User
Settings
Factory
Defaults
User
Settings
Fetal Rate Alert On/Off Off
High Rate Limit 160-210 bpm 180 bpm
Low Rate Limit 70-130 bpm 84 bpm
FECG Waveform On/Off On
FECG Waveform Size 0.10-2.00 1.00
FECG Sweep Speed 25 or 50 mm/sec 25 mm/sec
Twin Shift Auto/Manual/No shift No shift
Toco Sensitivity High/Medium/Low Medium
ECG Plot On/Off On
ECG Edit On/Off Off
External Device None/Nellcor FSpO
2
None
Print Period 1- to 60-minutes 10-minutes
Chart Recorder On/Off On
Speed 1, 2, or 3 cm/minute 3 cm/minute
Heat 1/2/3 2
Alignment 0-10 3
Scale USA/INTL USA
Variability Plot On/Off On
FHR1 Audio Level Off, 50% - 100% 75%
FHR2 Audio Level Off, 50% - 100% 75%
Patient Name Source Auto/Manual Auto
Time/Date Format USA/INTL USA
CIM Protocol A/B A
CIM Baud Rate 9.6k/19.2k 9.6k
Modem ID 0-255 0
Simulator On/Off Off
Begin Address 0-127 0
End Address 0-127 0
Increment 1-63 0
System Setup Key Visible/Invisible Invisible
Patient Admit/Discharge Save/Restore Restore
To review or change settings:
1 Touch FETAL.
2 Touch SETUP.
3 Touch the hidden key 3 times.
4 Touch DEFAULT STORAGE.
5 Touch USER SETTINGS.
6 Use CURSOR and CURSOR
to scroll through the parameter
description list.
7 Use and to change the
settings.
8 Touch NEXT PAGE to continue
to next screen of parameters.
9 Touch STORE to save the new
settings.
10 Touch YES when the Yes/No
message is displayed.