User's Manual

Bedside/Transport Monitors
5-21
Ultraview 1030/1050 with Capnography
(Option G or H)
Option G adds mainstream EtCO
2
monitoring to the Ultraview 1030/1050.
Figure 5-14: Ultraview 1030/1050
rear panel with option G or H
To start capnography
monitoring:
1 Plug the mainstream EtCO
2
sensor cable into the EtCO
2
receptacle on the capnography
panel.
2 Perform sensor calibration, if
necessary.
3 Prepare the patient according
to hospital procedures.
4 Select the appropriate airway
adapter (neonate or adult).
5 Verify that the windows are
clean and dry.
6 Place the sensor head over the
airway adapter and perform an
adapter calibration, if
necessary.
7 Remove the airway adapter
from the sensor head.
8 Insert the airway adapter into
the ventilator circuit and Ballard
style tracheal suction system (if
present) as shown.
9 Attach the sensor head to the
airway adapter.
10 Make the other connections as
shown.
11 Ensure that the sensor head is
always positioned above the
ventilator circuit so that
moisture will not enter the
adapter.
O
2
EtCO
2
endotracheal
mainstream
airway
ventilator
humidifier
adapter
O
2
sensor
capnography
panel
inspiratory circuit
expiratory circuit
O
2
adapter
cable
O
2
airway
adapter
tube
cable
sensor
head
sensor
ventilator
circuit