User's Manual

Bedside/Transport Monitors
5-15
Recordings are automatically inhibited when the battery power reaches
1/2 battery charge remaining.
The entire battery gauge flashes to emphasize this danger warning.
Wireless Network Interface
The Wireless Network Interface (WNI) provides Ultraview 1030/1050 monitors
equipped with the wireless option to communicate with the Spacelabs Medical
Ethernet network. This enables the wireless bedside monitor to communicate with
a central station and other networked devices.
The WNI provides wireless transmission of up to five waveforms along with
associated numerics, alarms, trends, arrhythmia, ST, and Flexport Interface
systems data.
Alarm Watch and Remote Trends are not restricted. Bed-to-bed communications
enable a bedside to use the Remote View or Alarm Watch features to view data
from other beds on the wireless or hardwired network.
Outbound Parameters Menu
To view the Outbound Parameters menu, touch the OUTBOUND PARAMETERS
key in the Monitor Setup menu.
When a parameter key is ON, the parameter is available for selection in another
monitor’s Remote View menu and Select Outbound Parameters menu. If two
parameters have identical labels, both parameters will display. The maximum
number of parameters that can be set depends upon the monitor’s specific
configuration for wireless networking. When the maximum number of parameters
is reached, the remaining parameters that were not selected become unavailable.
The monitor’s default outbound parameter selection is ECG1 when turned ON.
All parameters, including the parameters that are set to OFF in this menu, are
displayed by other monitors that are set to ALARM WATCH this monitor. If an
outbound parameter is set to OFF, all existing remote views for that parameter will
be dropped.
!
When the battery power reaches 1/2 battery charge remaining,
recordings that drain large amounts of power can cause the
monitor to power OFF without warning.
!
When the battery power becomes critically low (approximately
1/4 battery charge remaining), the monitor may power OFF at
any time, depending on monitor loading.
Setting OTHER to ON selects any unlisted channel types such
as VARI, CO, and unlabeled pressures.
To select parameters to be
monitored:
1 Touch MONITOR SETUP.
2 Touch OUTBOUND
PARAMETERS.
3 Select the parameters you wish
to display (up to five).
To change parameters:
1 Touch MONITOR SETUP.
2 Touch OUTBOUND
PARAMETERS.
3 Deselect the parameter key
that you wish to stop
displaying.
4 Select the parameter you wish
to display.