User's Manual

Bedside/Transport Monitors
5-13
Ultraview 1030/1050 Batteries
Each transport monitor is shipped with one 12 V battery. Monitors will accept two.
Batteries may be changed during battery or AC operation without losing data,
provided that one charged battery remains connected at all times during the
exchange.
A green LED in the lower-right corner of the front panel will flash when the
batteries are charging and will be continually illuminated once the batteries are
fully charged.
Sealed Lead Acid (SLA)
Sealed Lead Acid batteries only need to be charged for about 4-hours prior to use.
When storing, be sure to maintain their charge or to “top charge” them every three
months. These batteries are compatible with all 1030/1050 monitors.
Nickel Metal Hydride (NiMH)
Prior to use, Nickel Metal Hydride batteries should be run through three
charge/discharge cycles in the monitor. This will ensure that during the first
transport maximum run time will be possible. No other maintenance is required for
this battery type.
NiMH batteries may only be used in the Ultraview 1030/1050, with serial numbers
starting at 369-1xxxxx or 367-1xxxxx.
Figure 5-8: Ultraview 1030/1050 monitor battery installation
Ultraview 1030/1050
battery
battery
compartment
door
power on/off switch
LED