User's Manual

Bedside/Transport Monitors
5-11
Once the data is completely transferred, the monitor sounds a tone and displays
the message DATA TRANSFER COMPLETED below the UPDATE TRANSFER
and CANCEL TRANSFER keys. The module is now ready to be removed from
that monitor.
Figure 5-6: Data Transfer Completed display
If you do not transfer the module to the receiving monitor immediately, the source
monitor will be storing new data, but the module’s transfer data is not
automatically updated.
After 2-minutes, a tone will sound once each minute to remind you to update the
module with the newly acquired data and the monitor will display the message
DATA TRANSFER COMPLETED.
The message DATA IS XX MINUTES OLD displays the amount of time that has
elapsed since you last transferred the data. The monitor updates this message
each minute.
Updating the Data Transfer
If time has elapsed between data transfer into the module and module removal,
you may wish to update the transferred data before you remove the module. To
update this data, touch UPDATE TRANSFER.
While the data is being updated, the monitor re-displays the message
TRANSFERRING PATIENT DATA INTO MODULE.
Canceling the Data Transfer Before Module Removal
Touch the CANCEL TRANSFER key to clear all data transfer messages and keys
from the ECG display zone and purge previously transferred data from the
module. The monitor will then reactivate the TRANSFER DATA key in the ECG
Setup menu.
!
• Parameter information acquired between the time you transfer
data and the time you remove the module is lost. The monitor
indicates this lost information as a gap in the trends of the
parameters involved when you retrieve the data.
• To provide seamless trend information, remove the module and
insert it into the receiving monitor within 30- to 45-seconds after
you transfer the data. If there is a delay in completing the data
shuttle, update the transfer data just before you remove the
module.
• If the module has been out of the monitor for 10 or more
minutes, all data will be lost.
E
C
G
CANCEL TRANSFER
UPDATE TRANSFER
DATA TRANSFER COMPLETED
To update transfer data, touch
UPDATE TRANSFER.
To cancel data transfer, touch
CANCEL TRANSFER.