User's Manual

8-21
ECG Troubleshooting Guide
Clinical Situation Possible Cause Solution
AC noise Display resolution set to extended mode
(0.05 to 150 Hz).
Select monitor mode (0.5 - 40 Hz).
Electrodes dry. Repeat skin preparation and apply new
moist electrodes.
Patient cable entwined with other
electrical devices.
Separate patient cable from all other cables.
Baseline wanders Patient moving excessively. Use stress loops to secure lead wires and
cable to the patient.
Respiration artifact. Select another lead or reposition the
electrodes.
Electrodes dry. Repeat skin preparation and apply new
moist electrodes.
Low amplitude ECG Skin improperly prepared. Abrade skin and reapply electrodes.
Lead selected not showing QRS
complex with greatest amplitude.
Check 12-lead ECG to determine better
monitoring lead and reposition electrodes.
Electrodes could be positioned too near
bone or muscle mass.
Select another lead or reposition electrodes.
Module won’t learn ECG signal too noisy for initialization. Improve signal quality by repeating skin
preparation and/or repositioning electrodes.
ECG voltage below threshold. ECG
VOLTAGE TOO LOW message may be
displayed.
Perform the following steps as needed.
1. Check cables, lead wires and
electrodes, then relearn patient rhythm.
2. Change lead or reposition electrodes.
No ECG trace Improper attachment of ECG connector
cable to the module/or leads off.
Remove, then re-plug the connector into the
module or reconnect the leads.
LEDs flashing on front panel of module. Call a qualified service person.
Module is not seated into the monitor or
remote housing.
Remove, then reinsert the module or
exchange the module.
Excessive alarms Electrodes dry. Repeat skin preparation and apply new
moist electrodes.
Alarm limits set too close to patient’s
normal heart rate.
Readjust alarm limit.
Excessive interference: patient cable or
wires routed too close to other electrical
devices.
Reroute cables and leads.
Excessive patient movement or muscle
tremor.
Reposition electrodes and use stress loops
to secure lead wires and cable to the
patient.