User's Manual

Ultraview Care Network
7-32
Table 18: SpO
2
Alarm Attributes
Alarm Type
Tone Type Alarm Recording Alarm Watch
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
High Limit H,M,L,N H Y,N Y Y,N Y
Low Limit H,M,L,N H Y,N Y Y,N Y
Adapter
Disconnected
H,M,L,N L N N N Y,N N
Sensor
Disconnected
H,M,L,N L N N N Y,N N
Sensor Off
Patient
H,M,L,N L N N N Y,N N
Insufficient
Signal
H,M,L,N L N N N Y,N N
Light
Interference
H,M,L,N L N N N Y,N N
Noisy Signal H,M,L,N L N N N Y,N N
Low Signal
Strength
H,M,L,N N N N N Y,N N
H = High; M = Medium; L = Low; N = None; Y = Yes; N = No
!
• Setting tone type to NONE automatically sets alarm recording
and alarm watch to NO for that alarm. Changing tone type from
NONE to any other setting returns alarm recording and alarm
watch to factory defaults.
• Ultraview Digital Telemetry SpO
2
default settings are described
in Setting Up SpO2 Monitoring on page 14-4.