User's Manual

Module Configuration Manager
7-31
SpO
2
Default settings can be adjusted for high and low SpO
2
alarm limits and settings
for alarm priority.
Table 17: SpO
2
Parameter Configuration
Parameter
Available User
Settings
Factory
Defaults
User
Settings
SpO
2
Alarms On On
High Alarm limit 51 to 100% 100%
Low Alarm limit 50 to 99% 85%
Limit Alarm Delay 0- to 30-seconds 15-sec.
Message Alarm Delay 0- to 60-seconds 20-sec.
Data Averaging Time 4-, 8-, or 16-seconds 8-sec.
Pulse Tone (local only) On/Off Off
Pulse Tone Volume
(local only)
1 (soft) to 8 (loud) 4
Pulse Rate Display On/Off On
Motion Detection On/Off On
To review or change SpO
2
settings:
Module Configurations with ECG
1 Touch ECG.
2 Touch SETUP.
3 Touch CONFIG.
4 Touch the hidden key 3 times.
5 Touch USER SETTINGS.
6 Select SPO2.
7 Use CURSOR and CURSOR
to scroll through the parameter
description list.
8 Use and to change the
settings.
9 Touch NEXT PAGE to continue
to next screen of parameters.
10 Touch STORE to save the new
settings.
Module Configurations without
ECG
1 Touch TEMP.
2 Refer to steps 4 though 10
above.