User's Manual

Ultraview Care Network
7-30
Table 16: Invasive Pressure Alarm Attributes
Alarm Type
Tone Type Alarm Recording Alarm Watch
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
CH1 Systolic H,M,L,N H Y,N Y Y,N Y
CH1 Diastolic H,M,L,N H Y,N Y Y,N Y
CH1 Mean H,M,L,N H Y,N Y Y,N Y
CH1 CCP H,M,L,N H Y,N Y Y,N Y
CH2 Systolic H,M,L,N H Y,N Y Y,N Y
CH2 Diastolic H,M,L,N H Y,N Y Y,N Y
CH2 Mean H,M,L,N H Y,N Y Y,N Y
CH2 CCP H,M,L,N H Y,N Y Y,N Y
CH3 Systolic H,M,L,N H Y,N Y Y,N Y
CH3 Diastolic H,M,L,N H Y,N Y Y,N Y
CH3 Mean H,M,L,N H Y,N Y Y,N Y
CH3 CCP H,M,L,N H Y,N Y Y,N Y
CH4 Systolic H,M,L,N H Y,N Y Y,N Y
CH4 Diastolic H,M,L,N H Y,N Y Y,N Y
CH4 Mean H,M,L,N H Y,N Y Y,N Y
CH4 CCP H,M,L,N H Y,N Y Y,N Y
H = High; M = Medium; L = Low; N = None; Y = Yes; N = No
!
• Setting the tone type to NONE automatically sets the alarm
recording and alarm watch to NO for that alarm. Changing the
tone type from NONE to any other setting returns the alarm
recording and alarm watch to the factory defaults.