User's Manual

Ultraview Care Network
7-28
Table 14: Invasive Pressure Parameter Configuration (Channel 3)
Parameter Available User Settings Factory Defaults
User
Settings
Label (P3)
I C P, L A P, R A P, C V P, PA , A R T,
UA, UV, or PRS
PRS
Scales On/Off/Auto Off
Top of Scale 10 to 500 mmHg 200 mmHg
Filter Frequency
3, 8, 12, 15, 18, 21, 25, 30, 35, 40
Hz
12 Hz
Sweep Speed
50, 25, 12.5, 6.25 mm/sec., or
same as ECG
Same as ECG
Artifact Rejection On/Off Off
Numeric Size
Systolic/Diastolic Large
Mean Large
All Large
Systolic/Diastolic
Large
Alarm Type Learned/Fixed Learned
Systolic Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 180 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 100 mmHg
Diastolic Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 120 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 60 mmHg
Mean Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 130 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 80 mmHg
CPP Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 50 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 0 mmHg