User's Manual

Ultraview Care Network
7-26
Invasive Pressure
Invasive pressure parameters can be independently configured for each of two or
four pressure channels. The alarm type must be set to Fixed for the user-defined
alarm limit settings to be functional for high and low pressure alarm limits. If the
alarm type is set to the Learned mode, high and low alarm limits are set according
to the patient’s learned pressure.
The factory default label is:
ART for the first pressure channel label
PA for the second pressure channel
PRS for the third channel
CVP for the fourth channel
The associated alarm limit defaults correspond to typical pressure values for the
respective catheter site.
Table 12: Pressure Parameter Configuration Table (Channel 1)
Parameter Available User Settings Factory Defaults
User
Settings
Label (P1)
I C P, L A P, R A P, C V P, PA ,
ART, UA, UV, PRS
ART
Scales On/Off Off
Top of Scale 10 to 500 mmHg or Auto 200 mmHg
Filter Frequency
3, 8, 12, 15, 18, 21, 25,
30, 35,
40 Hz
12 Hz
Sweep Speed
50, 25, 12.5, 6.25
mm/sec., or same as ECG
same as ECG
Artifact Rejection On/Off Off
Numeric Size
Sys/Dia Large
Mean Large
All Large
Sys/Dia
Large
Alarm Type Learned/Fixed Learned
Systolic Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 180 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 100 mmHg
Diastolic Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 120 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 60 mmHg
Mean Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 130 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 80 mmHg
CPP Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 50 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 0 mmHg
To review or change Invasive
Pressure settings:
Module Configurations with ECG
1 Touch ECG.
2 Touch SETUP.
3 Touch CONFIG.
4 Touch the hidden key 3 times.
5 Touch USER SETTINGS.
6 Select PRESSURE.
7 Use CURSOR and CURSOR
to scroll through the parameter
description list.
8 Use and to change the
settings.
9 Touch NEXT PAGE to continue
to next screen of parameters.
10 Touch STORE to save the new
settings.
11 Touch YES when the STORE
ALL CHANNEL DEFAULTS is
displayed.
Module Configurations without
ECG
1 Touch TEMP.
2 Refer to steps 4 though 10
above.