Temperature
Default settings for high and low temperature alarm limits are adjustable, as well
as the settings for the delta temperature.
Table 10: Temperature Parameter Configuration
Parameter
Available User
Settings
Factory Defaults
User
Settings
Temperature 1 Alarms On/Off Off
High Alarm Limit 0.1° - 50.0°C 37.5°C
Low Alarm Limit 0.0° - 49.9°C 36.5°C
Temperature 2 Alarms On/Off Off
High Alarm Limit 0.1° - 50.0°C 37.5°C
Low Alarm Limit 0.0° - 49.9°C 36.5°C
Delta Alarms On/Off Off
High Alarm Limit 0.1° - 50.0°C 2.0°C
Low Alarm Limit 0.0° - 49.9°C 0.0°C
Default Storage Key Visible/Invisible Invisible
Patient
Admit/Discharge
Save/Restore Restore
Table 11: Temperature Alarm Attributes
Alarm Type
Tone Type Alarm Recording Alarm Watch
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
T1 High Limit H,M,L,N L Y,N N Y,N N
T1 Low Limit H,M,L,N L Y,N N Y,N N
T2 High Limit H,M,L,N L Y,N N Y,N N
T2 Low Limit H,M,L,N L Y,N N Y,N N
DT High Limit H,M,L,N L Y,N N Y,N N
DT Low Limit H,M,L,N L Y,N N Y,N N
H = High; M = Medium; L = Low; N = None; Y = Yes; N = No
!
• Setting the tone type to NONE automatically sets the alarm
recording and alarm watch to NO for that alarm. Changing the
tone type from NONE to any other setting returns the alarm
recording and alarm watch to the factory defaults.
To review or change
Temperature settings:
Module Configurations with ECG
1 Touch ECG.
2 Touch SETUP.
3 Touch CONFIG.
4 Touch the hidden key 3 times.
5 Touch USER SETTINGS.
6 Select TEMP.
7 Use CURSOR↑ and
CURSOR↓ to scroll through the
parameter description list.
8 Use ↑ and ↓ to change the
settings.
9 Touch STORE to save the new
settings.
10 Touch YES when the STORE
ALL CHANNEL DEFAULTS
message is displayed.
Module Configurations without
ECG
1 Touch TEMP.
2 Refer to steps 4 through 10
above.