User's Manual

Module Configuration Manager
7-23
Non-Invasive Blood Pressure
Non-invasive blood pressure (NIBP) alarms can be set independently for the adult
and infant modes. Default settings for systolic, diastolic and mean alarm limits are
fully adjustable within the specified ranges.
To have measurements automatically initiated at regular intervals, you can select
from the list of AUTO INTERVALS. After connecting a cuff to the patient, set
AUTO to ON in the NIBP menu and NIBP measurements will be automatic.
Table 8: NIBP Parameter Configurations
Parameter
Available User
Settings
Factory Defaults
User
Settings
Systolic Alarm (Adult) On/Off Off
High Alarm Limit 45 to 260 mmHg 150 mmHg
Low Alarm Limit 40 to 255 mmHg 100 mmHg
Diastolic Alarm (Adult) On/Off Off
High Alarm Limit 45 to 260 mmHg 100 mmHg
Low Alarm Limit 40 to 255 mmHg 60 mmHg
Mean Alarm (Adult) On/Off Off
High Alarm Limit 45 to 260 mmHg 110 mmHg
Low Alarm Limit 40 to 255 mmHg 90 mmHg
Systolic Alarm (Infant) On/Off Off
High Alarm Limit 15 to 150 mmHg 85 mmHg
Low Alarm Limit 10 to 145 mmHg 55 mmHg
Diastolic Alarm (Infant) On/Off Off
High Alarm Limit 15 to 150 mmHg 75 mmHg
Low Alarm Limit 10 to 145 mmHg 45 mmHg
Mean Alarm (Infant) On/Off Off
High Alarm Limit 15 to 150 mmHg 80 mmHg
Low Alarm Limit 10 to 145 mmHg 50 mmHg
NIBP Mode Adult/Neonate Adult
Display Pulse Rate Yes/No Yes
Auto Interval
1- to 10-minutes by 1
15- to 55-minutes by 5
1-, 2-, 4-, or 8-hours
15-minutes
To review or change NIBP
settings:
1 Touch ECG.
2 Touch SETUP.
3 Touch CONFIG.
4 Touch the hidden key 3 times.
5 Touch USER SETTINGS.
6 Select NIBP.
7 Use CURSOR and CURSOR
to scroll through the parameter
description list.
8 Use and to change the
settings.
9 Touch NEXT PAGE to continue
to next screen of parameters.
10 Touch STORE to save the new
settings.
11 Touch YES when the STORE
ALL CHANNEL DEFAULTS
message is displayed.