User's Manual

Module Configuration Manager
7-21
Varitrend
®
3
Default settings and individual parameter scales for the Varitrend timebase and
display are adjustable. Definitions for up to five events can be pre-selected to
describe concurrent changes in heart rate, respiration rate, or saturation level that
may be of specific clinical interest.
Table 7: Varitrend 3 Parameter Configuration
Parameter Available User Settings
Factory
Defaults
User
Settings
Trend Timebase 1.5 min. or 3.0 min. 1.5 min.
Resp Display Rate or waveform Waveform
Resp Waveform Size
+/- 0.3 ohms
+/- 0.5 ohms
+/- 1.0 ohms
+/- 2.0 ohms
+/- 6.0 ohms
+/- 2.0 ohms
Resp Rate Scale
AUTOSCALE
0 to 50
0 to100
0 to150
0 to 200
AUTOSCALE
Heart Rate Scale
AUTOSCALE
0 to100
0 to150
50 to150
0 to 200
100 to 200
0 to 250
100 to 250
0 to 300
100 to 300
AUTOSCALE
SpO
2
Scale
AUTOSCALE
0 to100
25 to100
50 to100
75 to100
AUTOSCALE
Auto Alarm
Recording
On/Off Off
Bradycardia Limit
(event 1)
0 to150 BPM or Off Off
Tachycardia Limit
(event 1)
100 to 300 BPM or Off Off
Apnea Limit (event 1) 10- to 40-seconds or Off Off
SpO
2
Limit (event 1) 25 to 95% or Off Off
Bradycardia Limit
(event 2)
0 to150 BPM or Off Off
Tachycardia Limit
(event 2)
100 to 300 BPM or Off Off
Apnea Limit (event 2) 10- to 40-seconds or Off Off
To review or change Varitrend 3
settings:
1 Touch ECG.
2 Touch SETUP.
3 Touch CONFIG.
4 Touch the hidden key 3 times.
5 Touch USER SETTINGS.
6 Select VARITREND.
7 Use CURSOR↑ and CURSOR↓
to scroll through the parameter
description list.
8 Use ↑ and ↓ to change the
settings.
9 Touch NEXT PAGE to continue
to next screen of parameters.
10 Touch STORE to save the new
settings.
11 Touch YES when the STORE
ALL CHANNEL DEFAULTS
message is displayed.