Table 6: Respiration Alarm Attributes
Alarm Type
Tone Type Alarm Recording Alarm Watch
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
Apnea H H H Y Y Y Y Y Y
High Rate H,M,L,N M Y,N Y Y,N Y
Low Rate H,M,L,N M Y,N Y Y,N Y
Loss of
Signal
H,M,L,N M N N N Y,N Y
H = High; M = Medium; L = Low; N = None; Y = Yes; N = No
!
• Alarms attributes for apnea cannot be changed.
- If RESP alarms are enabled, episodes of apnea will always
trigger a high alarm tone.
- If RESP alarm recording is enabled, apnea will always
trigger a recording.
- If Alarm Watch is enabled for a patient, apnea will always
trigger an alarm notification.
• Setting tone type to NONE automatically sets alarm recording
and alarm watch to NO for that alarm. Changing tone type from
NONE to any other setting returns alarm recording and alarm
watch to factory defaults.
• Setting tone type to NONE for high rate or low rate alarms
causes the affected alarm limit to be displayed in reverse video
in the RESP display zone.