User's Manual

Ultraview Care Network
7-18
NIBP Mean High Limit H,M,L,N L Y,N N Y,N N
NIBP Mean Low Limit H,M,L,N L Y,N N Y,N N
H = High; M = Medium; L = Low; N = None; Y = Yes; N = No
Messages and alarms beginning with “SPO2” or “NIBP” apply
only when the Ultraview Multiparameter Digital Telemetry
transmitter is used and multiparameter operation is enabled by
the MCM Multiparameter Telemetry ON setting. Refer to
Multiparameter Telemetry Overview on page 17-5 for more
information.
Alarm attributes for asystole and ventricular fibrillation cannot
be changed.
- If ECG alarms are enabled, episodes of asystole and
ventricular fibrillation will always trigger a high alarm tone.
- If ECG alarm recording is enabled, asystole and ventricular
fibrillation will always trigger a recording.
- If Alarm Watch is enabled for a patient, asystole and
ventricular fibrillation will always trigger an alarm
notification.
Setting tone type to NONE automatically sets alarm recording
and alarm watch to NO for that alarm. Changing tone type from
NONE to any other setting returns alarm recording and alarm
watch to factory defaults.
Setting tone type to NONE for high rate, low rate, ABN/MIN, or
run causes the affected alarm limit to be displayed in reverse
video in the ECG display zone.
Table 4: ECG Alarm Attributes (Telemetry Only) (continued)
Alarm Type
Tone Type Alarm Recording Alarm Watch
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings