User's Manual

Module Configuration Manager
7-15
Control (controls ALARM ON/OFF
and SUSPEND PROCESSING
keys)
On/Off On
Automatic Lead Switch On/Off On
Single Lead Alarm On/Off Off
Default Storage Key Visible/Invisible Invisible
Low Battery Alarm On/Off On
Patient Record On/Off Off
Telemetry Bed Assignment Temporary/Permanent Temporary
Patient Admit/Discharge Save/Restore Restore
Frequency Band
North America UHF, North
America VHF, European
Telecom, United Kingdom,
Australia, Spain, Sweden,
Finland, New Zealand, Norway,
Singapore, South Africa
North America
UHF
Primary Heart Rate Source ECG/ART/SPO2/UA ECG
Alternate Rate Source (ECG) On/Off On
Alternate Rate Source (ART) On/Off On
Alternate Rate Source (SPO2) On/Off On
Alternate Rate Source (UA) On/Off On
Multiparameter Telemetry On/Off On
Telemetry SpO2 Alarms
(only if multiparameter is On)
On On On
SpO2 High Limit
(only if multiparameter is On)
51% - 100% 100%
SpO2 Low Limit
(only if multiparameter is On)
50% - 99% 85%
SpO2 Limit Alarm Delay
(only if multiparameter is On)
0- to 30-seconds 15-seconds
SpO2 Message Alarm Delay
(only if multiparameter is On)
0- to 60-seconds 20-seconds
Telemetry NIBP Alarms
(only if multiparameter is On)
On On On
NIBP Systolic High Limit
(only if multiparameter is On)
35 to 260 mmHg 150 mmHg
NIBP Systolic Low Limit
(only if multiparameter is On)
30 to 255 mmHg 100 mmHg
NIBP Diastolic High Limit
(only if multiparameter is On)
35 to 260 mmHg 100 mmHg
Table 3: ECG Parameter Configuration (Telemetry Only) (continued)
Parameter Available User Settings
Factory
Defaults
User
Settings