User's Manual

Ultraview Care Network
7-14
QRS Tone (local only) On/Off Off
QRS Volume (local only) 1 (soft) to 8 (loud) 4
SpO
2
Pitch Modulation of QRS Tone On/Off Off
Display Resolution Monitor/Extended Monitor
Arrhythmia Processing On/Off On
Abnormals in a Row 2 to 10 or Off 5
Abnormals/Minute 1 to 99 or Off Off
Review On/Off On
1st Lead Selection Limb, V, augmented or MCL lead VI
1st Lead Waveform Size 2.00 mV/cm - 0.10 mV/cm 0.5 mV/cm
2nd Lead Selection Limb, V, augmented or MCL lead II
2nd Lead Waveform Size 2.00 mV/cm - 0.10 mV/cm 0.5 mV/cm
ST Level Alarm (channel 1) On/Off Off
ST Level Alarm Type Learned/Fixed Learned
High Level Alarm Limit
(only applies to fixed limits)
-8.75 mm to 9.00 mm 1.00 mm
Low Level Alarm Limit
(only applies to fixed limits)
-9.00 mm to 8.75 mm - 1.00 mm
High Level Offset
(only applies to learned limits)
0.25 mm to 9.00 mm 1.00 mm
Low Level Offset
(only applies to learned limits)
0.25 mm to 9.00 mm 1.00 mm
ST Change Alarm (channel 1) On/Off Off
Max-Min. Change Within 5-Minutes
(delta)
0.25 mm to 9.00 mm 1.00 mm
ST Level Alarm (channel 2) On/Off Off
ST Level Alarm Type Learned/Fixed Learned
High Level Alarm Limit
(only applies to fixed limits)
-8.75 mm to 9.00 mm 1.00 mm
Low Level Alarm Limit
(only applies to fixed limits)
-9.00 mm to 8.75 mm - 1.00 mm
High Level Offset
(only applies to learned limits)
0.25 mm to 9.00 mm 1.00 mm
Low Level Offset
(only applies to learned limits)
0.25 mm to 9.00 mm 1.00 mm
ST Change Alarm (channel 2) On/Off Off
Max-Min. Change Within 5-Minutes
(delta)
0.25 mm to 9.00 mm 1.00 mm
Check Setup Key On/Off On
Line Frequency 50 Hz, 60 Hz 60 Hz
Table 3: ECG Parameter Configuration (Telemetry Only) (continued)
Parameter Available User Settings
Factory
Defaults
User
Settings