User's Manual

Ultraview Care Network
7-12
Table 2: ECG Alarm Attributes
Alarm Type
Tone Type Alarm Recording Alarm Watch
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
Available User Settings
Factory Defaults
User Settings
Asystole H H H Y Y Y Y Y Y
VFIB H H H Y Y Y Y Y Y
High Rate H,M,L,N H Y,N Y Y,N Y
Low Rate H,M,L,N H Y,N Y Y,N Y
Run H,M,L,N H Y,N Y Y,N Y
Couplet H,M,L,N M Y,N Y Y,N Y
Ta c h H , M ,L,N M Y, N Y Y,N Y
ABN/MIN H,M,L,N M Y,N Y Y,N Y
ST Single-lead H,M,L,N M Y,N Y Y,N Y
ST Multi-lead H,M,L,N M Y,N Y Y,N Y
Classes Full H,M,L,N L N N N Y,N N
CH 1 & 2 Lead Off H,M,L,N M N N N Y,N Y
CH 1 Lead Off H,M,L,N M N N N Y,N Y
CH 2 Lead Off H,M,L,N M N N N Y,N Y
Low Voltage H,M,L,N M N N N Y,N Y
Noisy Signal H,M,L,N L N N N Y,N N
HR Unavailable H,M,L,N M N N N Y,N Y
H = High; M = Medium; L = Low; N = None; Y = Yes; N = No