User's Manual

Module Configuration Manager
7-9
For telemetry, some settings may or may not be available
depending on whether the module is used in a central or
bedside monitor.
For Ultraview Digital Telemetry (multiparameter), default
settings for SpO
2
and NIBP may be found as part of the ECG
channel. Telemetry ECG parameter configuration and alarm
attributes are found in separate tables in this chapter following
the tables for the hardwired (non-telemetry) products. Refer to
Table 1 and Table 2 for non-telemetry products. Refer to
Table 3 and Table 4 for telemetry products.
Table 1: ECG Parameter Configuration
Parameter Available User Settings
Factory
Defaults
User
Settings
ECG Rate Alarms (Adult)
Rate Alarm Type Learned or fixed limits Learned
High Rate Alarm Limit 5 to 300 BPM 180 BPM
Low Rate Alarm Limit 0 to 200 BPM 40 BPM
High Rate Offset
(only applies to learned limits)
5 to 80 BPM 50 BPM
Low Rate Offset
(only applies to learned limits)
5 to 80 BPM 50 BPM
ECG Rate Alarms (Infant)
Rate Alarm Type Learned or fixed limits Fixed
High Rate Alarm Limit 5 to 300 BPM 220 BPM
Low Rate Alarm Limit 0 to 200 BPM 100 BPM
High Rate Offset
(only applies to learned limits)
5 to 80 BPM 50 BPM
Low Rate Offset
(only applies to learned limits)
5 to 80 BPM 50 BPM
Infant Alarm Delay
(for ECG and Resp alarms)
0- to 3-minutes 3-minutes
ECG Mode Adult/Infant Adult
Sweep Speed 50, 25, or 12.5 mm/sec. 25 mm/sec.
Respiration On/Off On
Varitrend On/Off Off
QRS Tone (local only) On/Off Off
QRS Volume (local only) 1 (soft) to 8 (loud) 4
SpO
2
Pitch Modulation of
QRS Tone
On/Off Off
Display Resolution Monitor/Extended Monitor