User's Manual

Ultraview Care Network
6-8
Press NO to silence the alarms (if enabled), but to continue display of the DCI
notification. All alarms for that bedside monitor are silenced simultaneously. The
NO key becomes dithered (inactive) after it is pressed.
Enabling Alarms (Optional)
When activated, this option causes an alarm to sound and the parameter key to
flash when the DCI notification is displayed. This option must be enabled through
the Biomed menu.
A flashing parameter key and low level alarm tone, signifying an alarm condition,
will occur as soon as the central detects a communication loss. Alarm conditions
will continue until either YES or NO is pressed on the DCI notification, a new
parameter is assigned to the zone, or communication is restored.
Enabling Detection for Admitted Patients (Optional)
If ADMIT REQUIRED is ON, communication will be watched for only those remote
monitors having admitted patients. If ADMIT REQUIRED is OFF, communications
will be watched for all patients on the central monitor.