User's Manual

Ultraview Care Network
6-6
Setting the Time and Date
The TIME/DATE key controls both the time and date menus. The current time or
date is displayed above the menu.
Time is displayed in either 12- or 24-hour format. Network monitors display the
network time and stand-alone monitors display the internal system time.
The system will not make any time/date change permanent until you touch
ENTER.
System Clock
A system clock can be continually displayed in the lower right corner of the
screen. The clock feature can also be used as a stopwatch or timer.
Figure 6-3: Clock display formats
!
• Setting the time on any networked monitor sets time for all
monitors on that network.
!
• The digital clock is available for all monitors.
• The stopwatch and timer are only available for the UCW and
Ultraview 1700.
To change the network or internal
system time and date:
1 Touch MONITOR SETUP.
2 Touch PRIVILEGED ACCESS.
3 Enter Clinical password.
4 Touch TIME/DATE.
5a Select TIME.
-OR-
5b Select DATE.
6a Select 24-HOURS or AM/PM
(12-hour).
-OR-
6b Select MONTH, DAY, and
YEAR.
7 Use arrow keys to set time or
date.
8 Touch ENTER.
To enable/disable the clock:
1 Touch MONITOR SETUP.
2 Touch MONITOR
CONFIGURATION.
3a Select CLOCK ON to enable
the clock.
-OR-
3b Select CLOCK OFF to disable
the clock.
To access the clock menu:
1 Touch MONITOR SETUP.
2 Touch MONITOR
CONFIGURATION.
3 Select CLOCK ON.
4 Touch the clock that appears in
the lower right of the screen.
5 Select DIGITAL,
STOPWATCH, or TIMER.
digital
stopwatch
timer
02 JUN 03
02 JUN 03
02 JUN 03
Elapsed
Remaining