User's Manual
Table Of Contents
- Introduction
- About the Transmitters
- ECG and SpO2
- ECG Overview
- Patient Preparation and Electrode Application
- To Set Up ECG Monitoring
- ECG Problem Solving
- SpO2 Overview
- Warnings and Cautions for SpO2
- Setting Up SpO2 Monitoring
- Ensuring Accurate SpO2 Monitoring
- SpO2 and Pulse Rate Specifications
- Using the Sensorwatch Feature
- Enabling and Adjusting Alarms
- Data Averaging
- Display Details at the Host Monitor
- Printing SpO2 Waveforms
- SpO2 Messages at the Host Monitor
- SENSOR DISCONNECTED — Check connection at adapter cable
- SENSOR OFF PATIENT — Check connection at patient
- INSUFFICIENT SIGNAL — Reposition or replace sensor
- LOW SIGNAL STRENGTH — Reposition or replace sensor
- AMBIENT LIGHT INTERFERENCE — Cover sensor area
- NOISY SIGNAL
- FAULTY SENSOR — Replace sensor
- HARDWARE INCOMPATIBILITY — Contact service
- Sensors
- SpO2 Alarm Delays
- SpO2 Troubleshooting Guide
- Basic Operations
- Getting Started
- Basic Components
- Selecting Options for Leads
- Basic User Actions
- Basic Modes of Operation
- View Mode
- Status Messages at the Host Monitor
- Telemetry Transmitter with ECG Only Troubleshooting Guide
- Telemetry Transmitter with Display Troubleshooting Guide
- Telemetry Transmitter with Display and SpO2 Troubleshooting Guide
- Cleaning, Disinfecting, and Sterilization
- Appendix A — Guidance and Manufacturer’s Declaration
- Appendix B — Symbols
www.spacelabshealthcare.com
A-3
TELEMETRY TRANSMITTER (96281)
A
PPENDIX A — GUIDANCE AND
M ANUFACTURER’ S DECLARATION
Recommended
separation distance:
Where P is the
maximum output
power rating of the
transmitter in watts
(W), according to the
transmitter
manufacturer, and d is
the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey (a), should be
less than the
compliance level in
each frequency range.
Interference
may occur in the
vicinity of
equipment marked
with the following
symbol.
* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the monitoring device are used exceeds the applicable RF compliance level above, the 96281 telemetry
device should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the 96281 telemetry device.
Notes:
• At 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Table A-2 Electromagnetic Immunity
Immunity Test IEC 60601 Test Level
Compliance
Level
Electromagnetic
Environment
d =
1.17
P
80 MHz to 800 MHz
d =
2.34
P
800 MHz to 2.5 GHz