User's Manual
Table Of Contents
- Introduction
- About the Transmitters
- ECG and SpO2
- ECG Overview
- Patient Preparation and Electrode Application
- To Set Up ECG Monitoring
- ECG Problem Solving
- SpO2 Overview
- Warnings and Cautions for SpO2
- Setting Up SpO2 Monitoring
- Ensuring Accurate SpO2 Monitoring
- SpO2 and Pulse Rate Specifications
- Using the Sensorwatch Feature
- Enabling and Adjusting Alarms
- Data Averaging
- Display Details at the Host Monitor
- Printing SpO2 Waveforms
- SpO2 Messages at the Host Monitor
- SENSOR DISCONNECTED — Check connection at adapter cable
- SENSOR OFF PATIENT — Check connection at patient
- INSUFFICIENT SIGNAL — Reposition or replace sensor
- LOW SIGNAL STRENGTH — Reposition or replace sensor
- AMBIENT LIGHT INTERFERENCE — Cover sensor area
- NOISY SIGNAL
- FAULTY SENSOR — Replace sensor
- HARDWARE INCOMPATIBILITY — Contact service
- Sensors
- SpO2 Alarm Delays
- SpO2 Troubleshooting Guide
- Basic Operations
- Getting Started
- Basic Components
- Selecting Options for Leads
- Basic User Actions
- Basic Modes of Operation
- View Mode
- Status Messages at the Host Monitor
- Telemetry Transmitter with ECG Only Troubleshooting Guide
- Telemetry Transmitter with Display Troubleshooting Guide
- Telemetry Transmitter with Display and SpO2 Troubleshooting Guide
- Cleaning, Disinfecting, and Sterilization
- Appendix A — Guidance and Manufacturer’s Declaration
- Appendix B — Symbols
www.spacelabshealthcare.com
4-17
TELEMETRY TRANSMITTER (96281)
B
ASIC OPERATIONS
Telemetry Transmitter with Display and SpO
2
Troubleshooting Guide
Clinical Situation Possible Cause Solution
Low battery LED
flashing
• The batteries are low. • Replace batteries.
All leads off LEDs lit
• Self test failure. • Contact Spacelabs
Healthcare.
Any or all leads off
LEDs flashing
•Check ECG leads.
ECG is not shown at
Central Station
• Check if batteries are low. • Replace batteries.
SpO
2
data is not
shown on transmitter
display
•SpO
2
cable and/or sensor is
faulty.
• Unit malfunction.
•Check SpO
2
cable and
sensor.
• Contact Spacelabs
Healthcare.
SpO
2
is not shown at
Central Station
• Unit malfunction. • Contact Spacelabs
Healthcare.