User's Manual
Table Of Contents
- Introduction
- About the Transmitters
- ECG and SpO2
- ECG Overview
- Patient Preparation and Electrode Application
- To Set Up ECG Monitoring
- ECG Problem Solving
- SpO2 Overview
- Warnings and Cautions for SpO2
- Setting Up SpO2 Monitoring
- Ensuring Accurate SpO2 Monitoring
- SpO2 and Pulse Rate Specifications
- Using the Sensorwatch Feature
- Enabling and Adjusting Alarms
- Data Averaging
- Display Details at the Host Monitor
- Printing SpO2 Waveforms
- SpO2 Messages at the Host Monitor
- SENSOR DISCONNECTED — Check connection at adapter cable
- SENSOR OFF PATIENT — Check connection at patient
- INSUFFICIENT SIGNAL — Reposition or replace sensor
- LOW SIGNAL STRENGTH — Reposition or replace sensor
- AMBIENT LIGHT INTERFERENCE — Cover sensor area
- NOISY SIGNAL
- FAULTY SENSOR — Replace sensor
- HARDWARE INCOMPATIBILITY — Contact service
- Sensors
- SpO2 Alarm Delays
- SpO2 Troubleshooting Guide
- Basic Operations
- Getting Started
- Basic Components
- Selecting Options for Leads
- Basic User Actions
- Basic Modes of Operation
- View Mode
- Status Messages at the Host Monitor
- Telemetry Transmitter with ECG Only Troubleshooting Guide
- Telemetry Transmitter with Display Troubleshooting Guide
- Telemetry Transmitter with Display and SpO2 Troubleshooting Guide
- Cleaning, Disinfecting, and Sterilization
- Appendix A — Guidance and Manufacturer’s Declaration
- Appendix B — Symbols
www.spacelabshealthcare.com
3-23
TELEMETRY TRANSMITTER (96281)
ECG
AND SP O
2
No SpO
2
alarms are
shown
• ECG “Leads Off” condition
exists.
• Re-attach ECG leadwires to
the patient and resume ECG
monitoring.
• Higher priority alarm
condition is present.
• Clear the current alarm
condition and/or re-prioritize
SpO
2
alarms in the Module
Configuration Manager.
•When SpO
2
alarms are ON,
all SpO
2
alarm conditions
cause the parameter value to
blink according to the alarm
priority set using the Module
Configuration Manager.
FAULTY SENSOR
Replace Sensor
• A faulty or incompatible
sensor is connected to the
adapter cable.
• To clear the message,
connect a known good
sensor to the adapter cable.
• Sensor has failed.
Clinical Situation Possible Cause Solution