User's Manual
Table Of Contents
- Introduction
- About the Transmitters
- ECG and SpO2
- ECG Overview
- Patient Preparation and Electrode Application
- To Set Up ECG Monitoring
- ECG Problem Solving
- SpO2 Overview
- Warnings and Cautions for SpO2
- Setting Up SpO2 Monitoring
- Ensuring Accurate SpO2 Monitoring
- SpO2 and Pulse Rate Specifications
- Using the Sensorwatch Feature
- Enabling and Adjusting Alarms
- Data Averaging
- Display Details at the Host Monitor
- Printing SpO2 Waveforms
- SpO2 Messages at the Host Monitor
- SENSOR DISCONNECTED — Check connection at adapter cable
- SENSOR OFF PATIENT — Check connection at patient
- INSUFFICIENT SIGNAL — Reposition or replace sensor
- LOW SIGNAL STRENGTH — Reposition or replace sensor
- AMBIENT LIGHT INTERFERENCE — Cover sensor area
- NOISY SIGNAL
- FAULTY SENSOR — Replace sensor
- HARDWARE INCOMPATIBILITY — Contact service
- Sensors
- SpO2 Alarm Delays
- SpO2 Troubleshooting Guide
- Basic Operations
- Getting Started
- Basic Components
- Selecting Options for Leads
- Basic User Actions
- Basic Modes of Operation
- View Mode
- Status Messages at the Host Monitor
- Telemetry Transmitter with ECG Only Troubleshooting Guide
- Telemetry Transmitter with Display Troubleshooting Guide
- Telemetry Transmitter with Display and SpO2 Troubleshooting Guide
- Cleaning, Disinfecting, and Sterilization
- Appendix A — Guidance and Manufacturer’s Declaration
- Appendix B — Symbols
www.spacelabshealthcare.com
3-21
TELEMETRY TRANSMITTER (96281)
ECG
AND SP O
2
HARDWARE INCOMPATIBILITY — Contact service
The hardware configuration in your module is not compatible with
the programmed software options.
• Contact your Spacelabs Healthcare Field Service Engineer.
Sensors
For SpO
2
sensor compatibility, refer to the information that follows.
Spacelabs Healthcare SpO
2
Sensors
Please refer to the the accessories in Telemetry Transmitter (96281)
Data Sheet for compatible sensors (for the 96281-C) and adapter
cables.
Additional Information
For additional information about biocompatibility or sensor disposal,
refer to the manufacturer's instructions enclosed with each sensor.
SpO
2
Alarm Delays
SpO
2
Troubleshooting Guide
Table 3-6 SpO
2
Alarms
Alarm Setting Range Alarm Delay
Spacelabs Healthcare
Sat Low 50 to 99% • Less than (15 seconds + user selected alarm delay)
• Alarm delay is user selected from 0 to 30 seconds in
5 second increments.
Sat High 51 to 100% • Less than (20 seconds + user selected alarm delay)
• Alarm delay is user selected from 0 to 30 seconds in
5 second increments.
Clinical Situation Possible Cause Solution
No SpO
2
parameter
key in Channel Format
menu
•SpO
2
is not enabled at the
90478 receiver.
• Enable multiparameter
telemetry in the Module
Configuration Manager or
enable the SpO
2
display in
the Channel Format menu.