TELEMETRY TRANSMITTER (96281)
ECG
AND SP O
2
HARDWARE INCOMPATIBILITY — Contact service
The hardware configuration in your module is not compatible with
the programmed software options.
Contact your Spacelabs Healthcare Field Service Engineer.
Sensors
For SpO
2
sensor compatibility, refer to the information that follows.
Spacelabs Healthcare SpO
2
Sensors
Please refer to the the accessories in Telemetry Transmitter (96281)
Data Sheet for compatible sensors (for the 96281-C) and adapter
cables.
Additional Information
For additional information about biocompatibility or sensor disposal,
refer to the manufacturer's instructions enclosed with each sensor.
SpO
2
Alarm Delays
SpO
2
Troubleshooting Guide
Table 3-6 SpO
2
Alarms
Alarm Setting Range Alarm Delay
Spacelabs Healthcare
Sat Low 50 to 99% Less than (15 seconds + user selected alarm delay)
Alarm delay is user selected from 0 to 30 seconds in
5 second increments.
Sat High 51 to 100% Less than (20 seconds + user selected alarm delay)
Alarm delay is user selected from 0 to 30 seconds in
5 second increments.
Clinical Situation Possible Cause Solution
No SpO
2
parameter
key in Channel Format
menu
•SpO
2
is not enabled at the
90478 receiver.
Enable multiparameter
telemetry in the Module
Configuration Manager or
enable the SpO
2
display in
the Channel Format menu.