- SENSOR DISCONNECTED — Check connection at adapter cable
- SENSOR OFF PATIENT — Check connection at patient
- INSUFFICIENT SIGNAL — Reposition or replace sensor
- LOW SIGNAL STRENGTH — Reposition or replace sensor
- AMBIENT LIGHT INTERFERENCE — Cover sensor area
- NOISY SIGNAL
- FAULTY SENSOR — Replace sensor
- HARDWARE INCOMPATIBILITY — Contact service
- Sensors
- SpO2 Alarm Delays
- SpO2 Troubleshooting Guide
- Basic Operations
- Getting Started
- Basic Components
- Selecting Options for Leads
- Basic User Actions
- Basic Modes of Operation
- View Mode
- Status Messages at the Host Monitor
- Telemetry Transmitter with ECG Only Troubleshooting Guide
- Telemetry Transmitter with Display Troubleshooting Guide
- Telemetry Transmitter with Display and SpO2 Troubleshooting Guide
- Cleaning, Disinfecting, and Sterilization
- Appendix A — Guidance and Manufacturer’s Declaration
- Appendix B — Symbols
www.spacelabshealthcare.com
3-19
TELEMETRY TRANSMITTER (96281)
ECG
AND SP O
2
SENSOR DISCONNECTED — Check connection at adapter cable
This message and alarm indicate that the sensor is either
disconnected or the wiring is faulty.
• Check for proper sensor connection to the adapter cable.
• If the message persists, replace the sensor and/or the adapter
cable.
SENSOR OFF PATIENT — Check connection at patient
• The telemetry transmitter does not detect a valid sensor input
signal. Check the patient for proper sensor placement.
• The tissue between the LED and photodiode is too transmissive.
If the sensor placement seems correct and the message persists,
try a sensor site with a thicker tissue bed.
Note:
This message is not available with all SpO
2
sensors.
INSUFFICIENT SIGNAL — Reposition or replace sensor
• Insufficient signal for proper operation, indicated by a low
deflection on the Sensorwatch signal strength bar.
• Poor sensor application or site. Correctly reapply or reposition to
a better perfused site, or massage the site.
• If the message persists, then replace the sensor.
LOW SIGNAL STRENGTH — Reposition or replace sensor
When this message appears, the saturation and pulse rate continue
to show. However, the Sensorwatch bar flashes as an indication of a
possible error condition.
• Insufficient blood flow between the sensor light emitter and
detector. Move the sensor to an area of higher perfusion.
• Poor sensor application. Reposition to place active components
closer to the skin or locate to a better perfused site.
SPO2 SENSOR OFF PATIENT
SPO2 INSUFFICIENT SIGNAL
SPO2 AMBIENT LIGHT INTF.
SPO2 NOISY SIGNAL
SPO2 INSUFFICIENT SIGNAL
Table 3-5 Telemetry Message Equivalents (Continued)
Telemetry SpO
2
Message