User's Manual
Table Of Contents
- Introduction
- About the Transmitters
- ECG and SpO2
- ECG Overview
- Patient Preparation and Electrode Application
- To Set Up ECG Monitoring
- ECG Problem Solving
- SpO2 Overview
- Warnings and Cautions for SpO2
- Setting Up SpO2 Monitoring
- Ensuring Accurate SpO2 Monitoring
- SpO2 and Pulse Rate Specifications
- Using the Sensorwatch Feature
- Enabling and Adjusting Alarms
- Data Averaging
- Display Details at the Host Monitor
- Printing SpO2 Waveforms
- SpO2 Messages at the Host Monitor
- SENSOR DISCONNECTED — Check connection at adapter cable
- SENSOR OFF PATIENT — Check connection at patient
- INSUFFICIENT SIGNAL — Reposition or replace sensor
- LOW SIGNAL STRENGTH — Reposition or replace sensor
- AMBIENT LIGHT INTERFERENCE — Cover sensor area
- NOISY SIGNAL
- FAULTY SENSOR — Replace sensor
- HARDWARE INCOMPATIBILITY — Contact service
- Sensors
- SpO2 Alarm Delays
- SpO2 Troubleshooting Guide
- Basic Operations
- Getting Started
- Basic Components
- Selecting Options for Leads
- Basic User Actions
- Basic Modes of Operation
- View Mode
- Status Messages at the Host Monitor
- Telemetry Transmitter with ECG Only Troubleshooting Guide
- Telemetry Transmitter with Display Troubleshooting Guide
- Telemetry Transmitter with Display and SpO2 Troubleshooting Guide
- Cleaning, Disinfecting, and Sterilization
- Appendix A — Guidance and Manufacturer’s Declaration
- Appendix B — Symbols
www.spacelabshealthcare.com
3-11
TELEMETRY TRANSMITTER (96281)
ECG
AND SP O
2
Setting Up SpO
2
Monitoring
1 Connect the SpO
2
adapter cable to the patient-worn telemetry
transmitter.
2 Attach the sensor to the patient and connect the sensor cable to
the SpO
2
adapter cable.
3 Initiate ECG monitoring on the patient monitor.
4 Touch ECG.
5 Touch CHANNEL FORMAT.
6 Touch SpO
2
/ON.
Ensuring Accurate SpO
2
Monitoring
Each sensor requires site-specific application procedures. The
quality of the patient’s pulse oximetry measurements and pulse
signals may be adversely affected by certain environmental factors,
by oximetry sensor application errors, and by patient conditions.
Any of these factors can interfere with the monitor’s ability to detect
and show measurements and may result in a loss-of-pulse condition.
If the SpO
2
measurement does not seem reasonable, first check the
patient’s vital signs by alternate means and then check the pulse
oximeter for proper operation.
Patients with anemia and/or significant concentrations of
dysfunctional hemoglobins (such as carboxyhemoglobin,
methemoglobin, and sulphemoglobin) may appear to have normal
saturation values while actually being hypoxic. Further assessment,
using means other than pulse oximetry, is recommended for such
patients.
• For anemic patients, this condition occurs because patients have
decreased arterial oxygen contents.
• For patients with dysfunctional hemoglobins (that are unable to
carry oxygen), this condition occurs because less functional
hemoglobin is available to carry oxygen.
Other patient conditions that may result in inaccurate
measurements or a loss-of-signal condition during operation
include:
•Low perfusion
• Dark pigment
• Prolonged and/or excessive patient movement
Hemoglobin levels below 5 g/dl may prevent the monitor
from providing SpO
2
values.