TELEMETRY TRANSMITTER (96281)
A
BOUT THE TRANSMITTERS
Other combinations of disconnected electrodes yield no
useful data, resulting in no lead availability at either the
monitor or the transmitter.
Table 2-2 Lead Availability with Narrowband transmitters (96281-xxxNx)
Connected Electrodes (X)
Available leads at the
monitor
Shown leads
at the
transmitter
RA
(R)
LA
(L)
LL
(F)
RL
(N)
V
(C)
X X X X X II, V I, II, III, V
X X X X II, V I, II, III, V
X X X X I, II, III, aVR, aVF, aVL I, II, III
X X X I, II, III, aVR, aVF, aVL I, II, III
X X X X III III
X X X III III
X X X III III
X X III III
X X X X II II
XXXII II
XXXII II
X X II II
XX XXI I
XX X I I
XX XI I
XX I I
Table 2-3 Lead Availability with Wideband transmitters (96281-xxxWx)
Connected Electrodes (X)
Available leads at the
monitor
Shown leads
at the
transmitter
RA
(R)
LA
(L)
LL
(F)
RL
(N)
V
(C)
X X X X X I,II,III,V, aVF, aVL, aVR I, II, III, V
X X X X I,II,III, aVF, aVL, aVR I, II, III
X X X X I,II,III, aVF, aVL, aVR I, II, III
X X X X III III
X X X III III
X X X III III
X X X X II II
XXXII II