c. Troubleshooting with the ULTRAVIEW 1030 / ULTRAVIEW 1050 Site Survey screen.
NOTE: Current ULTRAVIEW-1030/ULTRAVIEW-1050 site survey screen does not provide
throughput information. Therefore this tool is only
to be used for network troubleshooting and
operational verifications.
To enter the SITE SURVEY mode:
Go to “MONITOR CONFIGURATION” Screen.
Select “PRIVILEGED ACCESS” button.
Enter Password.
Select “MORE” button.
Select “SERVICE FUNCTIONS” button.
Select “SYSTEM INFORMATION” button.
Select “SITE SURVEY” button.
ULTRAVIEW-1030/ULTRAVIEW-1050 site survey mode display provides important information for
verification of site operation.
Note
: In the ULTRAVIEW-1030/ULTRAVIEW-1050 Site survey, signal range is 80 to 230. A reading
below 100 should be considered as Marginal operation. A reading below 90 will experience dropouts and
should be cause for re-evaluation of WNI locations and coverage areas.
ULTRAVIEW-1030/ULTRAVIEW-1050 Site Survey Display
• The Channel number shown at the top of the screen is the current sync, or “Master”, channel number.
• In site survey the statistics screen is a running total of all nodes seen since the last system reset.
• Before each entry in the table there is either a “*” or “-“ indication. The “*” indicates node sent a
configuration message in the last 20 seconds. The “–“indicates configuration message not seen
within the last 20 seconds.
• Assure you are using an available node number for the survey monitor. If the node number of this bed
is in the survey table then the node number is a duplicate. Since the table only shows received
configuration packets it must never show a node with the same number as the survey tool node.
Part Number 062-0983-01 Rev C page 3 - 28