User's Manual

Again, back to our example. If the lowered throughput from WNI A is not attributable to transient causes
as listed above, WNI A may need to be relocated to an area where is will be out of range of the suspect
room. This will assure that a wireless monitor in the suspect room will synch to WNI B, where the
communications are better.
Remember that the effect of any change of WNI position must be evaluated from all rooms that can see
the WNI, not just the suspect room.
Part Number 062-0983-01 Rev C page 3 - 27